Influence of sample temperature and environmental humidity on measurements of benzene in ambient air by transportable GC-PID

relationships.isAuthorOfPublication
relationships.isSecondaryAuthorOf
relationships.isDirectorOf
Authors
Romero Trigueros, Cristina ; Doval Miñarro, Marta ; González Duperón, Esther ; González Ferradás, Enrique
item.page.secondaryauthor
item.page.director
Publisher
Copernicus Publications
publication.page.editor
publication.page.department
DOI
https://doi.org/10.5194/amt-10-4013-2017
item.page.type
info:eu-repo/semantics/article
Description
© Author(s) 2017. This manuscript version is made available under the CC-BY 4.0 license http://creativecommons.org/licenses/by/4.0/ This document is the Published version of a Published Work that appeared in final form in Atmospheric Measurement Techniques. To access the final edited and published work see https://doi.org/10.5194/amt-10-4013-2017, 2017
Abstract
Calibration of in situ analysers of air pollutants is usually done with dry standards. In this paper, the influence of sample temperature and environmental humidity on benzene measurements by gas chromatography coupled with a photoionisation detector (GC-PID) is studied. Two reference gas mixtures (40 and 5 µg m−3 nominal concentration benzene in air) were subjected to two temperature cycles (20/5/20 °C and 20/35/20 °C) and measured with two identical GC-PIDs. The change in sample temperature did not produce any significant change in readings. Regarding ambient humidity, the chromatographs were calibrated for benzene with dry gases and subjected to measure reference standards with humidity (20 and 80 % at 20 °C). When measuring a concentration of 0.5 µg m−3 benzene in air, the levels of humidity tested did not produce any significant interference in measurements taken with any of the analysers. However, when measuring a concentration of 40 µg m−3, biases in measurements of 18 and 21 % for each respective analyser were obtained when the relative humidity of the sample was 80 % at 20 °C. Further tests were carried out to study the nature of this interference. Results show that humidity interference depends on both the amount fractions of water vapour and benzene. If benzene concentrations in an area are close to its annual limit value (5 µg m−3), biases of 2.2 % can be expected when the absolute humidity is 8.6 g cm−3 – corresponding to a relative humidity of 50 % at 20 °C. This can be accounted for in the uncertainty budget of measurements with no need for corrections. If benzene concentrations are above the annual limit value, biases become higher. Thus, in these cases, actions should be taken to reduce the humidity interference, as an underestimation of benzene concentrations may cause a mismanagement of air quality in these situations.
publication.page.subject
Citation
Atmospheric Measurement Techniques, 2017, Vol. 10, Issue 10, pp. 4013-4022