Publication:
Optimización de las dosis de radiación en la artrografía de hombro

dc.contributor.authorCampos, P.A.
dc.contributor.authorRedondo, M.V.
dc.contributor.authorReus, M.
dc.contributor.authorMartínez Martínez, Francisco
dc.contributor.authorBerná Serna, Juan De Dios
dc.contributor.departmentCirugía, Pediatría y Obstetricia y Ginecología
dc.date.accessioned2025-05-08T11:39:39Z
dc.date.available2025-05-08T11:39:39Z
dc.date.issued2009-05-01
dc.description© 2008 SERAM. Publicado por Elsevier España, S.L. Todos los derechos reservados. This document is the Published version of a Published Work that appeared in final form in Radiología. To access the final edited and published work see https://doi.org/10.1016/j.rx.2008.11.006
dc.description.abstractObjetivo: El objetivo de este estudio fue reducir la dosis de radiación recibida por los pacientes sometidos a una artrografía de hombro y en los que se utiliza como sistema de guiado una placa con coordenadas radiopacas situada sobre el área de interés. Material y métodos: La dosis a la entrada se obtuvo en 34 pacientes con edades comprendidas entre 15 y 75 años, media de 44 años. La dosis a órganos de riesgo y la dosis efectiva se estimaron mediante técnicas de Monte Carlo, donde los parámetros de entrada son: anatomía del paciente, geometría de la exploración y kerma en aire a la entrada del paciente sin retrodispersión. Las artrografías se realizaron en un equipo telemando y las imágenes se obtuvieron mediante adquisición digital sin fluoroscopia. Resultados: El espesor medio de los hombros estudiados fue 14,6±2,1 cm (9–20 cm). Las imágenes se obtuvieron con 80±10 kVp (60–85 kVp) y 6,5±3,5 mAs (1,4–17 mAs). El tiempo medio de irradiación para cada paciente fue 20±6 ms (6,9–47,9 ms). El kerma en aire calculado fue de 0,41±0,19 mGy y la dosis efectiva de 0,79±0,40 μ Sv. Conclusiones: La técnica descrita en este trabajo ha permitido reducir la dosis de radiación al paciente respecto a otros procedimientos descritos en la bibliografía y que el radiólogo que realiza la artrografía no se irradie durante el procedimiento.es
dc.description.abstractObjective: The aim of this study was to determine whether using a film with radiopaque coordinates placed over the region of interest to guide shoulder arthrography can reduce the dose of radiation received by patients. Material and methods: The entrance dose was obtained in 34 patients (mean age, 44 years; range, 15 to 75 years). The dose received by organs at risk and the effective dose were estimated with Monte Carlo techniques using the following input parameters: patient anatomy, examination geometry, and air kerma at the entrance to the patient without backscattering. Arthrography was performed with a remote controlled device and images were acquired digitally without fluoroscopy. Results: The mean thickness of the shoulders studied was 14.6±2.1 cm (9–20 cm). Images were obtained with 80±10 kVp (60–85 kVp) and 6.5±3.5 mAs (1.4–17 mAs). The mean time of irradiation for each patient was 20±6 ms (6.9–47.9 ms). The calculated air kerma was 0.41±0.19 mGy and the effective dose was 0.79±0.40 μSv. Conclusions: The technique described in this study has enabled us to reduce the dose of radiation received by patients undergoing shoulder arthrography in comparison with other techniques described in the literature and to ensure that the radiologist performing the procedure is not irradiated.
dc.formatapplication/pdfes
dc.format.extent5es
dc.identifier.citationRadiología. 2009 51(3):282–286
dc.identifier.doihttps://doi.org/10.1016/j.rx.2008.11.006
dc.identifier.issnPrint: 0033-8338
dc.identifier.issnElectronic: 1578-1771
dc.identifier.urihttp://hdl.handle.net/10201/154277
dc.languagespaes
dc.publisherElsevier
dc.relationSin financiación externa a la Universidades
dc.relation.publisherversionhttps://www.sciencedirect.com/science/article/abs/pii/S0033833809000472?via%3Dihub
dc.rights.accessRightsinfo:eu-repo/semantics/restrictedAccess
dc.subjectArtrografíaes
dc.subjectHombroes
dc.subjectRadiometríaes
dc.subjectOptimizaciónes
dc.subjectProtección radiológicaes
dc.subjectArthrography
dc.subjectShoulder
dc.subjectRadiometry
dc.subjectOptimization
dc.subjectRadiological protection
dc.titleOptimización de las dosis de radiación en la artrografía de hombroes
dc.title.alternativeOptimization of the dose of radiation in shoulder arthrographyes
dc.typeinfo:eu-repo/semantics/articlees
dspace.entity.typePublicationes
relation.isAuthorOfPublicationb2db368b-5d8f-490c-bc1e-21b04b875f4f
relation.isAuthorOfPublication.latestForDiscoveryb2db368b-5d8f-490c-bc1e-21b04b875f4f
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
1-s2.0-S0033833809000472-main.pdf
Size:
224.05 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
2.26 KB
Format:
Item-specific license agreed upon to submission
Description:
Collections